Power
AC input:
100V-250V, 50Hz/60 Hz
AC power Consumption:
6.6W when battery is not being charged,
16W when battery is being charged
Operation
Working temperature:
10 -40°C (23 to 104°F)
Humidity:
30 -75% (non-condensing)
Atmospheric pressure:
86-106kPa
Battery Data
Type:
Lithium ion internal battery, rechargeable
Battery capacity:
12V,1800mAh, 8 hours (approximately)
Charging time:
4 hours from empty to full
Physical Data
Dimensions:
230mm(W)x160mm(H) x50mm(D)
Weight:
1150 grams
Capnography Measuring Data
EtCO2/CO2
Range:
0-99mmHg,or 0-13kPa,or 0-13%
Accuracy:
±2mmHg when 0-5%
±10%(reading) when > 5%
Update:
Once every breath
Trend length:
24hours
Respiration rate
Range:
3 -150 t/min
Accuracy:
the bigger value of ±10%(reading) or ±2 times/min
Trend length:
24 hours
PUMP Flow :
selectable from 50-250cc/min
Pulse Oximeter Data (optional)
Oximetry Saturation
Range:
0-99%
Accuracy:
±3% when 50-79%
±2% when 80-99%
Trend length:
24 hours
Pulse Rate
Range:
30-250bpm
Accuracy:
the bigger value of ±2bpm or ±5%(reading)
Trend length:
24 hours