Model:MP605
- Big screen LCD, clearly displays theEtCO2/CO2,Respiration rate,andSpO2,Pulse
(optional).
- EtCO2 Monitor(Capnograph) technology of infrared ray gas analysis:
1) Compensation system for temperature and pressure and Zero readjustment system
all these insure the accuracy and stability
2)Balance gas compensation makes minimum influence to other gases.
3)Lung pressure and temperature compensation (BTPS compensation) are selectable
4) Pump flux is selectable for adult and infant
- KPA , mmHg or % are selectable.
- Audio-optical alarm for exceeding limits and apnea
- 24 hours trend diagram keeps still available after power-off
- Side streammode, keeping work under the heavy humid environment, also suitable
for intubatton patient.
- Mainstreamprobe also can be used and more convenient.
- Data output via RS-232 port
- Portable and lithium battery
- DigitalSpO2system optional accuracy can be insured even under the movement or
weak perfusion.
- Easy to connect and fix with anesthetic machine.
Description
Power
AC input:
100V-250V, 50Hz/60 Hz
AC power Consumption:
6.6W when battery is not being charged,
16W when battery is being charged
Operation
Working temperature:
10 -40°C (23 to 104°F)
Humidity:
30 -75% (non-condensing)
Atmospheric pressure:
86-106kPa
Battery Data
Type:
Lithium ion internal battery, rechargeable
Battery capacity:
12V,1800mAh, 8 hours (approximately)
Charging time:
4 hours from empty to full
Physical Data
Dimensions:
230mm(W)x160mm(H) x50mm(D)
Weight:
1150 grams
Capnography Measuring Data
EtCO2/CO2
Range:
0-99mmHg,or 0-13kPa,or 0-13%
Accuracy:
±2mmHg when 0-5%
±10%(reading) when > 5%
Update:
Once every breath
Trend length:
24hours
Respiration rate
Range:
3 -150 t/min
Accuracy:
the bigger value of ±10%(reading) or ±2 times/min
Trend length:
24 hours
PUMP Flow :
selectable from 50-250cc/min
Pulse Oximeter Data (optional)
Oximetry Saturation
Range:
0-99%
Accuracy:
±3% when 50-79%
±2% when 80-99%
Trend length:
24 hours
Pulse Rate
Range:
30-250bpm
Accuracy:
the bigger value of ±2bpm or ±5%(reading)
Trend length: